Multifunction Tester, indication of touch voltage 25V or 50V measuring range overflow with blockade of measurement continuation, automatic circuit discharging after insulation resistance measurement and short circuit loop measurement without tripping of RCDs.
Continuity Testing | ||
---|---|---|
Measuring Range | 20/200Ω (Auto-Ranging) | |
Open Circuit Voltage | > 6V | |
Short Circuit Current | > 200mA | |
Accuracy | ±(3%rdg + 3dgt) | |
Insulation Resistance Testing | ||
Measuring Range | 20/200MΩ (Auto-Ranging) | |
Test Voltage | 500/1000V | |
Open Circuit Voltage | +20% - 0% | |
Rated Current | > 1mA | |
Accuracy | ± (3%rdg + 3dgt) | |
Loop Impedance Testing | ||
Impedance Range | 20/2000Ω | |
Rated Voltage | 230V + 10% - 15%(50Hz) | |
Normal Test Current | 20Ω:25A/10ms 2000Ω:15mA/350ms max. |
|
Accuracy | ±(3%rdg + 8dgt) | |
RCD Testing | ||
Test Current (Test Current Duration) |
x1/2, x1 | 10, 30, 100, 300, 500mA (2000ms) |
Fast | 150mA (50ms) | |
DC | 10, 30, 100, 300mA (2000ms), 500mA (200ms) | |
Auto Ramp | Goes up by 10% from 20% to 110% of IΔn. 300ms×10 | |
Rated Voltage | 230V + 10% - 15% 50Hz | |
Accuracy | Test Current | x1/2: -8%-2% x1, FAST: +2%+8% DC: ±10% Auto Ramp: ±4% |
Trip Time | ±(1%rdg + 3dgt) | |
Uc Testing | ||
Measuring Range | 100V | |
Rated Voltage | 230V + 10% - 15% (50Hz) | |
Test Current | 5mA at IΔn = 10mA 15mA at IΔn = 30/100mA 150mA at IΔn = 300/500mA |
|
Accuracy | +5% to +15%rdg±8dgt |
Safety Standard | IEC 61010-1 CAT III 300V Pollution Degree 2 IEC 61557-1, 2, 3, 4, 6, 10, IEC 60529 (IP40) |
---|---|
Power source | R6 or LR6 × 8 |
Dimensions | 115(L) × 175(W) × 86(D)mm |
Weight | 840g approx. |
Included Accessories | 7122B (Test leads) KAMP10* (Test Lead with IEC Connector) Pouch for Test leads Shoulder Strap Shoulder Pad Instruction Manual |
Optional Accessories | 7133B (Distribution board test leads) 8212-RS232C (RS232C adaptor with "KEW Report (Software)") 8212-USB (USB adaptor with "KEW Report (Software)") |